How often should a CMM be calibrated?
Most shops verify annually, tied to an internal procedure or a customer requirement. Heavier use, a harsh environment, or a recent crash justifies a shorter interval. The basis for the interval, not just the date, is what an auditor checks.
What is the difference between calibration and verification?
Calibration establishes the relationship between readings and reference values with a stated uncertainty. Performance verification confirms the machine still meets a stated specification, typically ISO 10360-2 or ASME B89.4.1. Most CMM work sold as calibration is verification.
What does ISO 10360-2 actually test?
Length measurement error, E0 and E150, is measured at five calibrated lengths, three times each, across seven positions including the four space diagonals: 105 measurements total, checked against the machine's stated MPE.
Can it be done on site, and how long is the machine down?
On-site verification is available across the Kansas City region and by scheduled trip farther out. Time on site depends on machine size and axis count and is quoted per job, not assumed from a price list.
What temperature does the room need to be?
ISO 1 sets 20 °C as the dimensional reference temperature. For an acceptance test the manufacturer's stated environment applies; for a reverification the customer's environment is recorded on the certificate, gradient included.
What happens if the machine fails the as-found test?
The as-found deviation is recorded before any adjustment. That data supports a reverse-traceability impact assessment: which parts were accepted on the machine, and whether the error could have changed an accept/reject decision.
Is there CMM calibration near me if my shop is outside Kansas City?
On-site routes extend into Lawrence, Topeka, St. Joseph and Ottawa on a regional schedule. Springfield and Wichita are farther out and scheduled as dedicated trips. Ship-in to the accredited laboratory remains available regardless of distance.
How do I check whether a certificate is accredited?
The certificate states the issuing laboratory and the standard applied. The laboratory's scope of accreditation, published by its accreditation body, lists the specific measurement, range and uncertainty covered — check the scope, not the logo on the certificate.
What is a CMC and why does it matter to my tolerances?
Calibration and measurement capability is the smallest uncertainty a laboratory can claim within its accredited scope. If the CMC does not comfortably support the tolerance being inspected, a 4:1 test uncertainty ratio is not achievable on that machine.
What is an error map, and can it be updated on my controller?
The volumetric compensation map corrects geometric error across the measuring volume. It is updated during geometric adjustment where the controller permits access; some OEM controllers lock that map to factory service only.